• Skip to navigation (Press Enter).
  • Skip to main content (Press Enter).
  • Universitätsklinikum
  • Home
  • Ophthalmology
    • Staff
    • Experimental Ophthalmology
    • Section Visual Function
    • Glaucoma
    • Histology lab
    • Corneal diseases
    • Hornhautbank
    • Cataract surgery
    • LASIK
    • Retinal dystrophies
    • Plastic surgery
    • Retinology
    • Neuro- and pediatric ophthalmology, strabism
    • Low vision aids
    • Uveitis
    • Hospital history
  • for Patients
  • for Ophthalmologists
  • for Students
  • Applications
  • Research, clinical studies
  • Events
  • Intranet (local only)
  • “our” man in Tansania
  • Friends of this department
  • Augennetz Südbaden
  • emergency clinic
Grafik unten
Logo
  • Contact
  • How to Find Us
  • Organisations
  • Media
Namensbildlinkesbildmittebildrechtesbild
Printer  Pdf
» Ophthalmology » Staff » » ·On-line miscellanea » ··DTL electrode
Google-Suchmaschine
Gesamte Uniklinikseite
Interne Suchmaschine
Aktuelle Uniklinikseite (augenklinik)
Suche nach Stichworten
Begriffe, Krankheiten, Einrichtungen u.v.m., verknüpft mit dem zuständigen Bereich.
Personensuche
Nachname:
Vorname:
Akad. Titel:
Org.-Einheit:
phonetische Suche
Suche nach Einrichtungen:
Such-Bereich:
Bezeichnung:
phonetische Suche
Suche über die Einrichtungen
Im Menüpunkt "Übersicht" sind Einrichtungen unter verschiedenen Gesichtspunkten gruppiert: Kliniken, Abteilungen, Institute, Zentrale Einrichtungen und ähnliches.
Smart-Link
Seitenaufruf mit Smartlink--> Info hier!
smartlink-symbol

→deutsche Version

Preparation and Montage of DTL™-like Electrodes
Michael Bach

Notes added 2010

• We mourn William W Dawson, one of the three inventors of the DTL™ electrode.

• The trademark for the DTL™ is held by Diagnosys LLC, and the electrodes can be obtained from them. I thank Bruce Doran for graciously allowing me to refer to the trademarked acronym.

 

Introductory Remark

I am describing here one possibility to mount DTL-like electrodes. This is not meant to be the last word on it, but its the current state after a decade of experience in this lab.

We nearly never use local anaesthetic. The slight foreign-body feeling wanes through the experiment, and local anaesthetic incurs the danger that the patient may hurt themselves, since the pain reflex is shut down.

We started using DTL fibres for the pattern ERG, but now also use it for luminance ERG. The electrode is inserted before dark adaptation, and the patient is counselled not to rub it out of the eye during the waiting period.

What is the DTL Electrode?

We should be thankful to the carpeting industry that they tackled the electrostatic charge problems that can arise with carpets made from plastic fibres. A metal-coated fibre was developed, which removes charges when weaved into the carpet due to its good electrical conducting properties. Just by chance this fibre is also well suited to record retinal potentials from the cornea. This was first discovered by Dawson, Trick & Litzkow in 1979, hence the name. Advantages of DTL-electrodes include:

  • DTL fibre is very flexible, giving little rise to irritation on the cornea, no local anaesthetic necessary
  • DTL fibre has a low electrical resistance (a few hundred Ohms)
  • DTL fibre, bought at the right place, is very inexpensive
  • DTL fibre can be discarded after each examination, so no disinfection is necessary afterwards.
DTL Montage, Principle

The very details of how you mount the electrode is less important than to do it identically every time. The figure shows too loose electrode insertion on the right; here it is likely that the fibre will roam around after blinks. Ideally, the fibre should not be so taut to cause irritation, but taut enough to avoid deep submersion in the lower lid.

There is a school of thought along the lines: “make it loose, it will dive deep down, but will thus be more comparable across subjects”. Quite possible, but I'm not aware of a paper demonstrating this.

Electrode Preparation

  • With two micro crocodile clips (from any electronics supplier) you catch a length of DTL fibre, sufficient for two eyes.
  • With an electrical continuity tester you check for continuous connection (even when shaking the clips), making sure that the coating of the DTL fibre is not broken, the clips are not corroded and the cables have no internal inkling of loosing connection.
  • Place the centre of the fibre on a sticky tape (width ~0.5 cm), whose sticky side is facing up.
  • Fix this with a counter-glued tiny tape.
  • Cut in the centre, and, lo, with one cut you get two electrodes (for the two eyes).
  • After recording the fibres can be discarded, avoiding infection and wear problems.
Electrode Preparation  

That's our spool with 80 kilometers of DTL-like fibre (maybe 79 km by now). As the manufacturer Statex tells us, this fibre called “Ex-Stat-Garn 22/1 dtex” consists of single-strand silver-coated nylon.
There is also multistrand fibre available, but the single strand allows for easier adjusment when held by the microclip.

Next to the spool the blue micro-crocodile clip can be seen with the hook protruding from its tip.
 

From the spool we take samples which are wrapped around cardboard and sterilised. Here a few turns of fibre can be seen – or not, as the fibre is very thin.
 

The floating end of the fibre is caught with a red micro-crocodile clip.
 

Caught the other end too, now cut an appropriate length (here experience is called for, I think 4 fingers are fine).
 

Now we place it on sticky tape (sunny side up). The spool only serves as holder for the sticky tape.
 

We press the fibre on the tape, and sandwich the fibre underneath another tiny strip of tape.
 

Now the electrode pair is fully taped.
 


With an electrical conductance tester –a simple ohmmeter with acoustic output– we check everything in one fell swoop: the grip of the clips, metal coating of the DTL fibre, and intactness of the electrode cables. The beep must not change when the entire contraption is moved, nor when careful pull is exerted on the clips. Then you know that conductivity is fine.
 


With a single snip we create two electrodes.
 
Always test the DTL conductivity! Occasionally there are “bad runs” in the spool – up to 1 m with intermittant contact problems, then it's fine again. Too bad should you discover this when it's in the eye already.
  
 
Here lies a pair of electrodes, and can't wait to be used...
 
The fibre is very thin indeed. There is only a hint of it to be seen, e.g. between the tape and directly above the black clip on the left or below the red clip on the right.


Mounting of DTL-like Electrodes

  1. The sticky tape at the end of the electrodes belongs nasally
  2. Look up! and the fibre is inserted into the eye.
  3. Temporally the tip of the micro croco clip is affixed.
  4. Because of the roundness of the check, the rear end of the croco clips has distance from the skin and is fixed in this position with a band of sticky tape.
  5. Done!
Mounting of DTL-like Electrodes  

Our subject already wears the temporal reference electrodes.
 

A blow-up up the reference electrode.
 

We start on the DTL by fixing the sticky tape nasally. Ideally, the tape should end up in a position that the fibre starts exactly in the middle of the lid angle.
 

Look up! Now the fibre is run temporally.
 

When placing the micro croco we find that the fibre is a bit long. No problem: loosen the grip of the croco clip and pull the end of the fibre slightly (not shown).
 

Tape above the croco clip.
 

The rear end of the croco clip has some distance from the skin, that's normal.
 

The floating end of the croco clip is held in place by a raised flap of transparent tape.
 

Voila a happy subject! The other eye also has an electrode in place. After placing a ground electrode anywhere (e.g. an earclip) the recording session can begin.

 

The first version of this page was published in April 1998.

©1998-2010 Prof. Michael Bach.